Q.

During an experiment with a metre bridge, the galvanometer shows a null point when the jockey is pressed at 40.0 cm using a standard resistance of 90 Ω, as shown in the figure. The least count of the scale used in the metre bridge is 1 mm. The unknown resistance is               [2014]

1 60±0.15Ω  
2 135±0.56Ω  
3 60±0.25Ω  
4 135±0.23Ω  

Ans.

(3)

In case of balanced meter bridge

Rl=X100-l

Given:    X=90 Ω,   l=40.0 cm

∴ R=Xl100-l=90×4060=60 Ω

∴ ΔRR=Δll+Δ(100-l)100-l⇒ΔR60=0.140+0.160

∴ ΔR=0.25 Ω

∴ R=(60±0.25) Ω